Autor(es): dc.contributor | Vianna, Gustavo André de Deus Carneiro | - |
Autor(es): dc.contributor | http://lattes.cnpq.br/8005271079083909 | - |
Autor(es): dc.contributor | http://lattes.cnpq.br/9424483240811213 | - |
Autor(es): dc.creator | Moreira, Edson Jorge Lima | - |
Data de aceite: dc.date.accessioned | 2024-07-11T18:34:49Z | - |
Data de disponibilização: dc.date.available | 2024-07-11T18:34:49Z | - |
Data de envio: dc.date.issued | 2023-05-17 | - |
Data de envio: dc.date.issued | 2023-05-17 | - |
Fonte completa do material: dc.identifier | http://app.uff.br/riuff/handle/1/28892 | - |
Fonte: dc.identifier.uri | http://educapes.capes.gov.br/handle/capes/773293 | - |
Descrição: dc.description | O presente trabalho teve duplo objetivo. O estudo 1 investigou os efeitos do uso clínico de instrumentos endodônticos reciprocantes M-Wire na rugosidade e na resistência à fratura cíclica e por torção dos instrumentos de NiTi. O estudo 2, isolando a variável tratamento térmico, investigou o desempenho mecânico de instrumentos de liga NiTi austenítica original e tratada termicamente. Metodologia. Estudo 1: Instrumentos Reciproc R25 foram testados em quatro situações diferentes, instrumentos novos como recebidos, após o preparo clínico de um, dois ou três molares superiores padronizados com quatro canais radiculares (n = 10 por grupo). O tempo total necessário para realizar cada preparo foi registrado, assim como o tempo para fratura por fadiga cíclica foram registrados. O teste de torção avaliou o torque e ângulo de rotação até a falha de acordo com a ISO 3630-1. A rugosidade das partes de trabalho dos instrumentos novos e usados foi avaliada com um perfilômetro (n = 5 por grupo). Análise estatística foi realizada usando ANOVA unidirecional e teste de Tukey a 5%. Estudo 2: Instrumentos ProFile foram submetidos ao tratamento térmico em forno com atmosfera de argônio a uma taxa de aquecimento de 10°C/min. Após atingir as temperaturas de 450ºC ou 500ºC, o sistema permaneceu em temperatura constante por 10 minutos, seguido de resfriamento em água à temperatura ambiente. Em seguida, os três grupos de instrumentos (n=30 por grupo) foram comparados quanto à fadiga cíclica (n=10 por grupo), flexão (n=10 por grupo) e resistência à flambagem (n=10 por grupo). A superfície e o modo de fratura foram observados em um microscópio eletrônico de varredura. A análise estatística foi realizada por meio dos testes de ANOVA unidirecional e teste de Student-Newman-Keuls a 5%. Resultados. Estudo 1 - Não foram observadas fraturas ou deformações após o uso clínico. O tempo de preparo foi maior para os instrumentos usados três vezes. Não houve diferenças significativas entre os grupos em relação à fadiga ou resistência à torção (p > 0,05). Com relação às medidas de rugosidade, a profundidade do sulco foi maior em instrumentos novos e com um uso do que nos instrumentos utilizados duas ou três vezes (p < 0,05). Estudo 2: Os instrumentos adquiriram coloração amarelada e azulada após tratamentos térmicos a 450 ºC e 500 ºC, respectivamente. Os instrumentos ProFile originais apresentaram menor resistência à fadiga cíclica e maior resistência à flexão e à flambagem (P<0.05). Em contraste, os instrumentos ProFile azulados (tratados termicamente a 500°C) apresentaram maior resistência à fadiga cíclica e menor resistência à flexão e à flambagem (P>0.05). Conclusões – Estudo 1 – quanto maior foi o número de vezes de utilização, maior foi o tempo para o preparo e menor foi a rugosidade da superfície do instrumento; no entanto, o uso clínico não afetou a fadiga cíclica ou a resistência à torção dos instrumentos reciprocantes. Estudo 2 – os diferentes tratamentos térmicos realizados nos instrumentos ProFile aumentaram sua resistência à fadiga cíclica e melhoraram a flexibilidade e resistência à flambagem. Palavras-chave: Instrumentos odontológicos; | - |
Descrição: dc.description | The present study had a dual objective. Study 1 investigated the effects of the clinical use of M-Wire reciprocating endodontic instruments on the roughness and resistance to cyclic fracture and torsion of NiTi instruments. Study 2, isolating a thermal variable, investigated the mechanical treatment of original and heat-treated austenitic NiTi alloy instruments. Methodology. Study 1: R25 instruments were tested in four different situations: new instruments, after one, two and three clinical uses on upper molar with four canals (n= 10 per group). The total time required to perform each preparation was recorded, as well as the time to cyclic fatigue were recorded. The torque and test angle with rotation up to a failure in accordance to ISO 3630-1. The Surface roughness of the working parts of the new instruments and those used once, twice and three times was evaluated by using a Profilometer (n= 5 per group). Statistical analysis was performed using one-way ANOVA and 5% Tukey's test. Study 2: ProFile instruments were selected for heat treatment carried out in a furnace in argon atmosphere using a heating rate of 10° C/min. After reaching the programmed temperatures of 450ºC or 500 ºC the system remained at a constant temperature for 10 minutes, followed by cooling in water at room temperature. Afterwards, the three groups (n=30 per group) of instruments were compared regarding their cyclic fatigue (n=10 per group), bending (n=10 per group), and buckling resistance (n=10 per group). After cyclic fatigue tests, a scanning electron microscope was used to analyze the fracture surfaces and observe the fracture mode. Statistical analysis was performed using One-way ANOVA and Student-Newman-Keuls test, with an alpha type error set at 0.05. Results. Study 1 - No fractures or deformations were observed after clinical use. Higher preparation time was needed during the third use of the instruments for all root canals (p < 0.05). There were no significant differences among the groups regarding either cyclic fatigue or torsional resistance (p > 0.05). Regarding the roughness measurements, groove depth was higher on new and one- versus two- or three-maxillary-molar-prepared instruments (p < 0.05). Study 2: Yellowish and blueish coloration was observed in the ProFile instruments after 450 ºC or 500 ºC heat treatments, respectively. Conventional ProFile instruments showed the lower cyclic fatigue, and the higher bending and buckling resistance (P<0.05). In contrast, yellowish ProFile instruments (heat treated at 500° C) showed the higher cyclic fatigue, and the lower bending and buckling resistance (P>0.05). Conclusions – Study 1: It can be concluded that the clinical use of Reciproc instruments increased preparation time and decreased surface roughness. However, clinical use did not affect the cyclic fatigue or torsional resistance of the Reciproc instruments. Study 2: It can be concluded that the different heat treatments performed on ProFile instruments increased its cyclic fatigue resistance and improved the flexibility and buckling resistance. | - |
Descrição: dc.description | 62 f. | - |
Formato: dc.format | application/pdf | - |
Idioma: dc.language | pt_BR | - |
Direitos: dc.rights | Open Access | - |
Direitos: dc.rights | CC-BY-SA | - |
Palavras-chave: dc.subject | Instrumentos odontológicos | - |
Palavras-chave: dc.subject | Endodontia | - |
Palavras-chave: dc.subject | Tratamento do canal radicular | - |
Palavras-chave: dc.subject | Instrumentos endodônticos | - |
Palavras-chave: dc.subject | NiTi | - |
Palavras-chave: dc.subject | Propriedades mecânicas | - |
Palavras-chave: dc.subject | Endodontia | - |
Palavras-chave: dc.subject | Tratamento do canal radicular | - |
Palavras-chave: dc.subject | Instrumento odontológico | - |
Palavras-chave: dc.subject | Fadiga de materiais | - |
Palavras-chave: dc.subject | Teste mecânico | - |
Palavras-chave: dc.subject | Teste de dureza | - |
Palavras-chave: dc.subject | Dental instruments | - |
Palavras-chave: dc.subject | Endodontics | - |
Palavras-chave: dc.subject | Root canal treatment | - |
Palavras-chave: dc.subject | Endodontic instruments | - |
Palavras-chave: dc.subject | NiTi | - |
Palavras-chave: dc.subject | Mechanical properties | - |
Título: dc.title | Uso clínico e desempenho mecânico de instrumentos de NiTi tratados termicamente considerando-se sua resistência à fadiga cíclica e por torção e rugosidade da superfície | - |
Título: dc.title | Clinical use and mechanical performance of heat-treated NiTi instruments considering their resistance to cyclic and torsional fatigue and surface roughness | - |
Tipo de arquivo: dc.type | Tese | - |
Aparece nas coleções: | Repositório Institucional da Universidade Federal Fluminense - RiUFF |
O Portal eduCAPES é oferecido ao usuário, condicionado à aceitação dos termos, condições e avisos contidos aqui e sem modificações. A CAPES poderá modificar o conteúdo ou formato deste site ou acabar com a sua operação ou suas ferramentas a seu critério único e sem aviso prévio. Ao acessar este portal, você, usuário pessoa física ou jurídica, se declara compreender e aceitar as condições aqui estabelecidas, da seguinte forma: