Torque de inserção, remoção e fratura de mini-implantes ortodônticos

Registro completo de metadados
MetadadosDescriçãoIdioma
Autor(es): dc.contributorMucha, José Nelson-
Autor(es): dc.contributorVilella, Oswaldo de Vasconcellos-
Autor(es): dc.contributorElias, Carlos Nelson-
Autor(es): dc.creatorLoss, Tatiana Féres Assad-
Data de aceite: dc.date.accessioned2024-07-11T17:59:05Z-
Data de disponibilização: dc.date.available2024-07-11T17:59:05Z-
Data de envio: dc.date.issued2018-08-27-
Data de envio: dc.date.issued2018-08-27-
Data de envio: dc.date.issued2009-
Fonte completa do material: dc.identifierhttps://app.uff.br/riuff/handle/1/7228-
Fonte: dc.identifier.urihttp://educapes.capes.gov.br/handle/capes/761365-
Descrição: dc.descriptionObjetivo: Avaliar a relação entre a forma e a resistência à fratura em torção de 5 diferentes mini-implantes ortodônticos. Material e Método: 25 mini-implantes divididos em 5 grupos (DEN, RMO, CON, NEO, SIN) foram submetidos ao ensaio de fratura em torção na região do pescoço e da ponta, através de mandris acoplados à Máquina Universal de Ensaios Mecânicos (EMIC, Curitiba, PR, Brasil). A forma da ponta ativa dos mini-implantes foi avaliada no Microscópio Óptico (Stemi 2000-C - Zeiss, Jena, Alemanha) e no MEV (JEOL, modelo JSM-5800 LV, Tóquio, Japão). A resistência da liga dos mini-implantes foi avaliada através de tensão cisalhante e da tensão normal a partir do valor de torque. Resultados: O torque de fratura do pescoço variou de 23,45 N.cm (DEN) a 34,82 N.cm (SIN) e o torque de fratura da ponta variou de 9,35 N.cm (CON) a 24,36 N.cm (NEO). As características que mais influenciaram os resultados foram: diâmetro externo, diâmetro interno, a proporção entre diâmetro interno e externo, e a existência do fresado na região apical do mini-implante. As tensões diferiram entre os grupos, havendo diferença entre as ligas de confecção dos mini implantes, entretanto não houve relação entre os valores das tensões e dos torques de fratura. Conclusão: Os torque de fratura foram diferentes tanto para o pescoço como para a ponta nos 5 tipos avaliados e foram influenciados pela forma dos mini-implantes. A resistência da liga de confecção dos mini-implantes não teve influência significativa nos valores do torque de fratura-
Descrição: dc.descriptionThis study evaluated the influence of the mini-implants‟ design on fracture resistance of torsion. Material and Method: 25 mini-implants divided into 5 groups (DEN, RMO, CON, NEO, SIN) were tested on the twisting of fracture in the neck and the tip, through a mandrel coupled in Universal Testing Machine Mechanical (EMIC, Curitiba, PR, Brazil). The design of the active edge of mini-implants were evaluated by optical microscope (STEM 2000-C - Zeiss, Jena, Germany) and SEM (JEOL, model JSM-5800 LV, Tokyo, Japan). The alloy resistance of mini-implants was evaluated by shear tension and normal tension using normal torque value as baseline. Results: The torque value needed for the neck fractures ranged from 23,45 N.cm (DEN) to 34,82 N.cm (SIN) and for the tip fractures ranged from 9.35 N.cm (CON) to 24.36 N.cm (NEO). The design characteristics that most influenced the results were: outside and inside diameter, the ratio between diameters, and the cutter. The tension values differ among the groups. There was difference among mini-implants alloy; however there was no relationship between the values of shear and normal tension and fracture torque. Conclusion: The mini-implants‟ design had an influence on the fracture resistance of torsion. Torques of fracture was different either for the neck and for the tip for all 5 groups evaluated. The mini-implants manufacture alloy resistance has no significant influence on the torque fracture values-
Formato: dc.formatapplication/pdf-
Idioma: dc.languagept_BR-
Direitos: dc.rightsopenAccess-
Direitos: dc.rightsopenAccess-
Direitos: dc.rightshttp://creativecommons.org/licenses/by-nc-nd/3.0/br/-
Direitos: dc.rightsCC-BY-SA-
Palavras-chave: dc.subjectMini-implante ortodôntico-
Palavras-chave: dc.subjectTorque de fratura-
Palavras-chave: dc.subjectForma-
Palavras-chave: dc.subjectOrtodontia-
Palavras-chave: dc.subjectProcedimento de ancoragem ortodôntica-
Palavras-chave: dc.subjectImplante dentário-
Palavras-chave: dc.subjectOrthodontic mini-implant-
Palavras-chave: dc.subjectFracture torque-
Palavras-chave: dc.subjectMini-implant design-
Título: dc.titleTorque de inserção, remoção e fratura de mini-implantes ortodônticos-