Determination of rare earth elements in Fe-minerals using external calibration by LA-ICP-MS and application on Cauê Iron Formation (Brazil).

Registro completo de metadados
MetadadosDescriçãoIdioma
Autor(es): dc.creatorSousa, Denise Versiane Monteiro de-
Autor(es): dc.creatorAbreu, Adriana Trópia de-
Autor(es): dc.creatorSampaio, Geraldo Magela Santos-
Autor(es): dc.creatorLana, Cristiano de Carvalho-
Autor(es): dc.creatorRodrigues, Daniel Aparecido da Silva-
Autor(es): dc.creatorNalini Júnior, Hermínio Arias-
Data de aceite: dc.date.accessioned2022-02-21T19:58:15Z-
Data de disponibilização: dc.date.available2022-02-21T19:58:15Z-
Data de envio: dc.date.issued2021-07-27-
Data de envio: dc.date.issued2021-07-27-
Data de envio: dc.date.issued2019-
Fonte completa do material: dc.identifierhttp://www.repositorio.ufop.br/jspui/handle/123456789/13419-
Fonte completa do material: dc.identifierhttps://www.sciencedirect.com/science/article/abs/pii/S0039914020305300-
Fonte completa do material: dc.identifierhttps://doi.org/10.1016/j.talanta.2020.121239-
Fonte: dc.identifier.urihttp://educapes.capes.gov.br/handle/capes/650404-
Descrição: dc.descriptionThe rare earth elements (REE) composition in Fe-mineral phases is an important tool in iron formation studies to obtain information about parent rocks and environmental and paragenetic processes. However, the determination of REE presents some difficulties, such as the low concentration of these elements, matrix complexity and lack of iron matrix certified reference materials. The aim of the present work is to propose an analytical method to determine the REE plus Y (REE + Y) contents at trace levels in Fe-(hydr)oxides by the laser ablation ICPquadrupoleMS technique, using external calibration. The calibration curves were obtained from analyses of reference materials with different matrices, and the analytical conditions were checked on the NIST 614 glass. The linearity (R2 ≥ 0.98), limit of detection (0.002–0.044 μg g−1), limit of quantification (0.008–0.146 μg g−1), recovery (88.4–112.4%), and intraday (0.1–14.1%) and interday (1.6–17.8%) precision were systematically assessed. The results obtained showed that the method is fit for the purpose and showed evidence of a nonsignificant interference of the matrix. Thus, the developed procedure was applied in the analyses of magnetite, martite, hematite, and goethite grains from Cauê Iron Formation (Brazil). The REE + Y patterns of the minerals are consistent with the previous study of bulk analyses on whole rocks and highlight the postdepositional signature of these elements in banded iron formations.-
Formato: dc.formatapplication/pdf-
Idioma: dc.languageen-
Direitos: dc.rightsrestrito-
Palavras-chave: dc.subjectREE+Y-
Palavras-chave: dc.subjectFe-(hydr)oxides-
Palavras-chave: dc.subjectBanded iron formations-
Título: dc.titleDetermination of rare earth elements in Fe-minerals using external calibration by LA-ICP-MS and application on Cauê Iron Formation (Brazil).-
Aparece nas coleções:Repositório Institucional - UFOP

Não existem arquivos associados a este item.