McGrath MAC Versus Three-Dimensional Printed Video Laryngoscopes: A Randomized, Manikin-Simulated Noninferiority Controlled Study with Medical Students

Registro completo de metadados
MetadadosDescriçãoIdioma
Autor(es): dc.contributorUniversidade Estadual Paulista (UNESP)-
Autor(es): dc.contributorSanta Casa de Misericórdia Bahia-
Autor(es): dc.contributorUniversidade Federal da Bahia (UFBA)-
Autor(es): dc.contributorD’Or Institute for Research and Education (IDOR)-
Autor(es): dc.contributorState University of Feira de Santana-
Autor(es): dc.creatorDetoni, Pablo B.-
Autor(es): dc.creatorNascimento, Jedson S.-
Autor(es): dc.creatorAraújo Azi, Liana M.T.-
Autor(es): dc.creatorPustilnik, Alexandre G.-
Autor(es): dc.creatorGusmão-Cunha, André-
Autor(es): dc.creatorMódolo, Norma Sueli P.-
Autor(es): dc.creatorCampos, Guilherme O.-
Autor(es): dc.creatorde Almeida, Victor S.-
Autor(es): dc.creatorCambui, João Pedro M.M.-
Autor(es): dc.creatorde Almeida, Vinicius S.-
Autor(es): dc.creatorAlves, Rodrigo L.-
Data de aceite: dc.date.accessioned2025-08-21T18:49:58Z-
Data de disponibilização: dc.date.available2025-08-21T18:49:58Z-
Data de envio: dc.date.issued2025-04-29-
Data de envio: dc.date.issued2025-01-31-
Fonte completa do material: dc.identifierhttp://dx.doi.org/10.1213/ANE.0000000000007067-
Fonte completa do material: dc.identifierhttps://hdl.handle.net/11449/303990-
Fonte: dc.identifier.urihttp://educapes.capes.gov.br/handle/11449/303990-
Descrição: dc.descriptionBACKGROUND: Proficiency in endotracheal intubation (ETI) is essential for medical professionals and its training should start at medical schools; however, large caseload may be required before achieving an acceptable success rate with direct laryngoscopy. Video laryngoscopy has proven to be an easier alternative for intubation with a faster learning curve, but its availability in medical training may be an issue due to its high market prices. We devised a low-cost 3-dimensionally printed video laryngoscope (3DVL) and performed a randomized trial to evaluate if the intubation success rate on the first attempt with this device is noninferior to a standard commercially available video laryngoscope (STVL). METHODS: Two hundred and nine medical students from 5 medical schools were enrolled and randomized to start with the STVL (McGrath MAC) or the 3DVL. Four stations (standard airway using the STVL/3DVL and difficult airway using the STVL/3DVL) were set to simulate ETI with standard Airway Management Trainer manikins (Laerdal Medical Ltd.). The noninferiority margin of 7.5% was defined for the success rate on the first attempt, considering the difference in proportions between the STVL (expected to be higher) and 3DVL groups. RESULTS: Regarding the standard airway station, 60.7% (n = 65) of the students successfully performed TI on the first attempt with the STVL within the established timeframe, compared to 36.3% (n = 37) of the students using the 3DVL. This represented a difference of 24.4% (95% confidence interval, 17.5%–31.3%). Considering the difficult airway station, the success rates on the first intubation attempt with the 2 VLs did not differ. CONCLUSIONS: The 3DVL was inferior in achieving first-attempt intubation when compared with the STVL with a difference in success rate >7.5% margin in simulated scenarios with medical students. Tracheal intubation might require a set of psychomotor skills for which the McGrath MAC device is superior to the low-cost alternative.-
Descrição: dc.descriptionDepartment of Surgical Specialties and Anesthesiology São Paulo State University (UNESP) Medical School-
Descrição: dc.descriptionDepartment of Anesthesiology Santa Casa de Misericórdia Bahia-
Descrição: dc.descriptionDepartment of Anesthesiology and Surgery Bahia School of Medicine Federal University of Bahia-
Descrição: dc.descriptionDepartment of Anesthesiology São Rafael Hospital D’Or Institute for Research and Education (IDOR)-
Descrição: dc.descriptionDepartment of Computer Engineering PGCC-UEFS State University of Feira de Santana-
Descrição: dc.descriptionDepartment of Surgical Specialties and Anesthesiology São Paulo State University (UNESP) Medical School-
Formato: dc.format334-341-
Idioma: dc.languageen-
Relação: dc.relationAnesthesia and Analgesia-
???dc.source???: dc.sourceScopus-
Título: dc.titleMcGrath MAC Versus Three-Dimensional Printed Video Laryngoscopes: A Randomized, Manikin-Simulated Noninferiority Controlled Study with Medical Students-
Tipo de arquivo: dc.typelivro digital-
Aparece nas coleções:Repositório Institucional - Unesp