In Vitro Quantitative Evaluation of Postprocessing Filter for Metal Artifact Reduction in Cone Beam Computed Tomography Images of Titanium and Zirconium Dioxide Implants

Registro completo de metadados
MetadadosDescriçãoIdioma
Autor(es): dc.contributorCruzeiro Do sul University (UNICSUL)-
Autor(es): dc.contributorUniversidade Estadual Paulista (UNESP)-
Autor(es): dc.contributorAnkara University-
Autor(es): dc.creatorCosta, Andre Luiz Ferreira-
Autor(es): dc.creatorFardim, Karolina A. C.-
Autor(es): dc.creatorMantoani, Jennifer M.-
Autor(es): dc.creatorRicardo, Ana Lucia Franco-
Autor(es): dc.creatorJardini, Maria Aparecida N.-
Autor(es): dc.creatorOrhan, Kaan-
Autor(es): dc.creatorDe Castro Lopes, Sérgio Lúcio Pereira-
Data de aceite: dc.date.accessioned2025-08-21T21:37:59Z-
Data de disponibilização: dc.date.available2025-08-21T21:37:59Z-
Data de envio: dc.date.issued2022-04-28-
Data de envio: dc.date.issued2022-04-28-
Data de envio: dc.date.issued2021-12-31-
Fonte completa do material: dc.identifierhttp://dx.doi.org/10.1155/2022/1362473-
Fonte completa do material: dc.identifierhttp://hdl.handle.net/11449/223681-
Fonte: dc.identifier.urihttp://educapes.capes.gov.br/handle/11449/223681-
Descrição: dc.descriptionObjective. To evaluate a postprocessing filter of a new imaging-processing software for analysis of metal artifact reduction. Methods. Eight artificial edentulous mandibles (phantoms), where titanium and zirconium dioxide implants had been installed in four different regions (i.e., incisors, canine, premolars, and molars). CBCT volume was acquired, and then, four types of filters were applied to the images: BAR filter and Multi-CDT NR filter (e-Vol DX) and Sharpening Filters 1x and 2x (OnDemand). Artifact was assessed by measuring the standard deviation (SD) of the gray values of filtered and unfiltered images. The comparison between implant material, teeth, and filters was performed by using ANOVA, whereas multiple comparisons were performed by using Bonferroni's test. The level of significance adopted was 5%. Results. The results showing higher SD values, which suggests a worse image, were obtained with titanium implants compared to zirconium dioxide ones. With regard to the four filters used, it can be seen that the lowest SD values were obtained with BAR and Multi-CDT NR filters and the highest with Sharpening Filters 1x and 2x, with no statistical difference between them, except regarding the molar region in titanium implants. Conclusion. The highest SD values were seen in zirconium dioxide implants, mainly in the region of anterior teeth. The BAR filter was found to be the most effective as its SD value decreased significantly, indicating that the image quality was improved.-
Descrição: dc.descriptionCruzeiro Do sul University (UNICSUL), SP-
Descrição: dc.descriptionDepartment of Diagnosis and Surgery Science and Technology Institute São Paulo State University, São José dos Campos-
Descrição: dc.descriptionDepartment of Dentomaxillofacial Radiology Faculty of Dentistry Ankara University-
Descrição: dc.descriptionDepartment of Diagnosis and Surgery Science and Technology Institute São Paulo State University, São José dos Campos-
Idioma: dc.languageen-
Relação: dc.relationBioMed Research International-
???dc.source???: dc.sourceScopus-
Título: dc.titleIn Vitro Quantitative Evaluation of Postprocessing Filter for Metal Artifact Reduction in Cone Beam Computed Tomography Images of Titanium and Zirconium Dioxide Implants-
Tipo de arquivo: dc.typelivro digital-
Aparece nas coleções:Repositório Institucional - Unesp

Não existem arquivos associados a este item.